4-2
Troubleshooting
Introduction
This section of the manual will assist you in troubleshooting and repairing the Model
2425. Included are self-tests, test procedures, troubleshooting tables, and circuit descrip-
tions. Note that disassembly instructions are located in Section 5, and component layout
drawings are at the end of Section 6.
Safety considerations
WARNING
The information in this section is intended for qualified service
personnel only. Do not perform these procedures unless you are
qualified to do so. Some of these procedures may expose you to
hazardous voltages that could cause personal injury or death. Use
caution when working with hazardous voltages.
Repair considerations
Before making any repairs to the Model 2425, be sure to read the following
considerations.
CAUTION
The PC-boards are built using surface mount techniques and require
specialized equipment and skills for repair. If you are not equipped
and/or qualified, it is strongly recommended that you send the unit
back to the factory for repairs or limit repairs to the PC-board
replacement level. Without proper equipment and training, you could
damage a PC-board beyond repair.
• Repairs will require various degrees of disassembly. However, it is recommended
that the Front Panel Tests be performed prior to any disassembly. The disassembly
instructions for the Model 2425 are contained in Section 5 of this manual.
• Do not make repairs to surface mount PC-boards unless equipped and qualified to
do so. (See previous CAUTION.)
• When working inside the unit and replacing parts, adhere to the handling
precautions and cleaning procedures explained in Section 5.
• Many CMOS devices are installed in the Model 2425. These static-sensitive
devices require special handling as explained in Section 5.
• Whenever a circuit board is removed or a component is replaced, the Model 2425
must be recalibrated. See Section 2 for details on calibrating the unit.
Содержание SourceMeter 2425
Страница 1: ...Model 2425100W SourceMeter Service Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031...
Страница 15: ......
Страница 16: ...1 PerformanceVerification...
Страница 35: ...1 20 Performance Verification Model 2425 Service Manual...
Страница 36: ...2 Calibration...
Страница 62: ...3 RoutineMaintenance...
Страница 65: ...3 4 Routine Maintenance Model 2425 Service Manual...
Страница 66: ...4 Troubleshooting...
Страница 84: ...5 Disassembly...
Страница 92: ...6 ReplacementParts...
Страница 113: ...6 22 Replacement Parts Model 2425 Service Manual...
Страница 114: ...A Specifications...
Страница 116: ...B CommandReference...
Страница 128: ...C CalibrationPrograms...
Страница 139: ......
Страница 141: ......