2-26
Calibration
Step 5: Save calibration constants.
Calibration is now complete, so you can store the calibration constants in EEROM by
sending the following command:
:CAL:PROT:SAVE
NOTE
Calibration will be temporary unless you send the SAVE command. Also,
calibration data will not be saved if (1) calibration is locked, (2) invalid data
exists, or (3) all steps were not completed.
Step 6: Lock out calibration.
To lock out further calibration, send the following command after completing the calibra-
tion procedure:
:CAL:PROT:LOCK
Single-range calibration
Normally, the complete calibration procedure should be performed to ensure that the
entire instrument is properly calibrated. In some instances, however, you may want to
calibrate only certain ranges. To do so, complete the entire procedure only for the range(s)
to be calibrated. Keep in mind, however, that you must complete all parameter steps for
each source or sense range. Also be sure to set calibration dates and save calibration after
calibrating the desired range(s).
Содержание SourceMeter 2425
Страница 1: ...Model 2425100W SourceMeter Service Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031...
Страница 15: ......
Страница 16: ...1 PerformanceVerification...
Страница 35: ...1 20 Performance Verification Model 2425 Service Manual...
Страница 36: ...2 Calibration...
Страница 62: ...3 RoutineMaintenance...
Страница 65: ...3 4 Routine Maintenance Model 2425 Service Manual...
Страница 66: ...4 Troubleshooting...
Страница 84: ...5 Disassembly...
Страница 92: ...6 ReplacementParts...
Страница 113: ...6 22 Replacement Parts Model 2425 Service Manual...
Страница 114: ...A Specifications...
Страница 116: ...B CommandReference...
Страница 128: ...C CalibrationPrograms...
Страница 139: ......
Страница 141: ......