Calibration
2-17
Step 4: Enter calibration dates and save calibration.
NOTE
For temporary calibration without saving new calibration constants, proceed to
Step 5: Lock out calibration.
1. From normal display, press MENU.
2. Select CAL, and then press ENTER. The Model 2425 will display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES
©
ß
SAVE LOCK CHANGE-PASSWORD
3. Select SAVE, then press ENTER. The unit will display the following message:
SAVE CAL
Press ENTER to continue; EXIT to abort calibration sequence
4. Press ENTER. The unit will prompt you for the calibration date:
CAL DATE: 12/15/2000
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
5. Change the displayed date to today's date, then press the ENTER key. Press
ENTER again to confirm the date.
6. The unit will then prompt for the calibration due date:
NEXT CAL: 12/15/2001
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
7. Set the calibration due date to the desired value, then press ENTER. Press ENTER
again to confirm the date.
8. Once the calibration dates are entered, calibration is complete, and the following
message will be displayed:
CALIBRATION COMPLETE
Press ENTER to confirm; EXIT to abort
9. Press ENTER to save the calibration data (or press EXIT to abort without saving
calibration data.) The following message will be displayed:
CALIBRATION SUCCESS
Press ENTER or EXIT to continue.
10. Press ENTER or EXIT to complete process.
Step 5: Lock out calibration.
1. From normal display, press MENU.
2. Select CAL, then press ENTER. The Model 2425 will display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES
©
ß
SAVE LOCK CHANGE-PASSWORD
3. Select LOCK, then press ENTER to lock out calibration.
Содержание SourceMeter 2425
Страница 1: ...Model 2425100W SourceMeter Service Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031...
Страница 15: ......
Страница 16: ...1 PerformanceVerification...
Страница 35: ...1 20 Performance Verification Model 2425 Service Manual...
Страница 36: ...2 Calibration...
Страница 62: ...3 RoutineMaintenance...
Страница 65: ...3 4 Routine Maintenance Model 2425 Service Manual...
Страница 66: ...4 Troubleshooting...
Страница 84: ...5 Disassembly...
Страница 92: ...6 ReplacementParts...
Страница 113: ...6 22 Replacement Parts Model 2425 Service Manual...
Страница 114: ...A Specifications...
Страница 116: ...B CommandReference...
Страница 128: ...C CalibrationPrograms...
Страница 139: ......
Страница 141: ......