Troubleshooting
4-5
Principles of operation
The following information is provided to support the troubleshooting tests and procedures
covered in this section of the manual. Refer to the following drawings:
Figure 4-1 — Overall block diagram
Figure 4-2 — Analog circuitry block diagram
Figure 4-3 — Power supply block diagram
Figure 4-4 — Output stage simplified schematic
Figure 4-5 — Digital circuitry block diagram
Overall block diagram
Figure 4-1 shows an overall block diagram of the Model 2425. Circuitry may be divided
into three general areas:
• Analog circuits — includes sourcing circuits such as the DACs, clamps, output
stage, and feedback circuits, as well as measurement circuits such as the A/D con-
verter.
• Digital circuits — includes the microcomputer that controls the analog section,
front panel, and GPIB and RS-232 ports, as well as associated interfacing circuits.
• Power supplies — converts the AC line voltage into DC voltages that supply the
power for the digital and analog circuits, and the output stage.
Содержание SourceMeter 2425
Страница 1: ...Model 2425100W SourceMeter Service Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031...
Страница 15: ......
Страница 16: ...1 PerformanceVerification...
Страница 35: ...1 20 Performance Verification Model 2425 Service Manual...
Страница 36: ...2 Calibration...
Страница 62: ...3 RoutineMaintenance...
Страница 65: ...3 4 Routine Maintenance Model 2425 Service Manual...
Страница 66: ...4 Troubleshooting...
Страница 84: ...5 Disassembly...
Страница 92: ...6 ReplacementParts...
Страница 113: ...6 22 Replacement Parts Model 2425 Service Manual...
Страница 114: ...A Specifications...
Страница 116: ...B CommandReference...
Страница 128: ...C CalibrationPrograms...
Страница 139: ......
Страница 141: ......