6. Select EXECUTE, then press ENTER. The instrument will display the following
message:
I-CAL
Press ENTER to 10.000
μ
A
7. Press ENTER. The Model 2425 will 10
μ
A and simultaneously display the
following:
DMM RDG: +10.00000
μ
A
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
8. Note and record the DMM reading, and then adjust the Model 2425 display to
agree exactly with the actual DMM reading. Use the up and down arrow keys to
select the digit value, and use the left and right arrow keys to choose the digit posi-
tion (or use the number keys, 0-9, +/-). Note that the display adjustment range is
within ±10% of the present range.
9. After adjusting the display to agree with the DMM reading, press ENTER. The
instrument will then display the following:
I-CAL
Press ENTER to 00.000
μ
A
10. Press ENTER. The Model 2425 will source 0mA and at the same time display the
following:
DMM RDG: +00.00000
μ
A
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
11. Note and record the DMM reading, and then adjust the Model 2425 display to
agree with the actual DMM reading. Note that the display value adjustment limits
are within ±1% of the present range.
12. After adjusting the display value to agree with the DMM reading, press ENTER.
The unit will then display the following:
I-CAL
Press ENTER to Output -10.000
μ
A
13. Press ENTER. The Model 2425 will source -10
μ
A and display the following:
DMM RDG: -10.00000
μ
A
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
14. Note and record the DMM reading, and then adjust the Model 2425 display to
agree with the DMM reading. Again, the maximum display adjustment is within
±10% of the present range.
15. After adjusting the display value to agree with the DMM reading, press ENTER
and note that the instrument displays:
I-CAL
Press ENTER to 00.000
μ
A
16. Press ENTER The Model 2425 will source 0
μ
A and simultaneously display the
following:
DMM RDG: +00.00000
μ
A
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
Содержание SourceMeter 2425
Страница 1: ...Model 2425100W SourceMeter Service Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031...
Страница 15: ......
Страница 16: ...1 PerformanceVerification...
Страница 35: ...1 20 Performance Verification Model 2425 Service Manual...
Страница 36: ...2 Calibration...
Страница 62: ...3 RoutineMaintenance...
Страница 65: ...3 4 Routine Maintenance Model 2425 Service Manual...
Страница 66: ...4 Troubleshooting...
Страница 84: ...5 Disassembly...
Страница 92: ...6 ReplacementParts...
Страница 113: ...6 22 Replacement Parts Model 2425 Service Manual...
Страница 114: ...A Specifications...
Страница 116: ...B CommandReference...
Страница 128: ...C CalibrationPrograms...
Страница 139: ......
Страница 141: ......