1-14
Performance Verification
5. Repeat the procedure for negative output currents with the same magnitudes as
those listed in Table 1-5.
6. Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
3A range accuracy
1. With the power off, connect the digital multimeter and the 1
Ω
resistor to the Model
2425 INPUT/OUTPUT jacks, as shown in Figure 1-3.
2. Select the multimeter DC volts measuring function.
3. Press the Model 2425 SOURCE I key to source current, and make sure the source
output is turned on.
Figure 1-3
Connections for 3A range current verification tests
2425 100 W SourceMeter
250V
PEAK
5V
PEAK
HI
LO
OUTPUT
125V
PEAK
125V
PEAK
EDIT
TOGGLE
POWER
RANGE
INPUT/
OUTPUT
4- WIRE
SENSE
DISPLAY
ON/OFF
TERMINALS
FRONT/
REAR
AUTO
RANGE
!
®
EXIT
ENTER
CONFIG MENU
SWEEP
TRIG
REL
LOCAL
FILTER
LIMIT
DIGITS SPEED
V
Ω
MEAS
I
FCTN
V
I
SOURCE
2
3
0
1
6
7
8
9
4
+/-
5
STORE
RECALL
EDIT
Digital Multimeter
Model 2425
Input HI
Input LO
1
Ω
Resistor
Содержание SourceMeter 2425
Страница 1: ...Model 2425100W SourceMeter Service Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031...
Страница 15: ......
Страница 16: ...1 PerformanceVerification...
Страница 35: ...1 20 Performance Verification Model 2425 Service Manual...
Страница 36: ...2 Calibration...
Страница 62: ...3 RoutineMaintenance...
Страница 65: ...3 4 Routine Maintenance Model 2425 Service Manual...
Страница 66: ...4 Troubleshooting...
Страница 84: ...5 Disassembly...
Страница 92: ...6 ReplacementParts...
Страница 113: ...6 22 Replacement Parts Model 2425 Service Manual...
Страница 114: ...A Specifications...
Страница 116: ...B CommandReference...
Страница 128: ...C CalibrationPrograms...
Страница 139: ......
Страница 141: ......