3. After sending a calibration command, repeatedly test the ESB (Event Summary)
bit (bit 5) in the status byte until it is set. (Use *STB? to request the status byte.)
4. Once operation complete has been detected, clear OPC status using one of two
methods: (1) use the *ESR? query, then read the response to clear the standard
event status register, or (2) send the *CLS command to clear the status registers.
Note that sending *CLS will also clear the error queue and operation complete
status.
Generating an SRQ on calibration complete
An IEEE-488 bus SRQ (service request) can be used to detect operation complete instead
of repeatedly polling the Model 2425. To use this method, send both *ESE 1 and *SRE 32
to the instrument, then include the *OPC command at the end of each calibration com-
mand line, as covered above. Clear the SRQ by querying the ESR (using the *ESR?
query) to clear OPC status, then request the status byte with the *STB? query.
Refer to your controller's documentation for information on detecting and servicing SRQs.
Содержание SourceMeter 2425
Страница 1: ...Model 2425100W SourceMeter Service Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031...
Страница 15: ......
Страница 16: ...1 PerformanceVerification...
Страница 35: ...1 20 Performance Verification Model 2425 Service Manual...
Страница 36: ...2 Calibration...
Страница 62: ...3 RoutineMaintenance...
Страница 65: ...3 4 Routine Maintenance Model 2425 Service Manual...
Страница 66: ...4 Troubleshooting...
Страница 84: ...5 Disassembly...
Страница 92: ...6 ReplacementParts...
Страница 113: ...6 22 Replacement Parts Model 2425 Service Manual...
Страница 114: ...A Specifications...
Страница 116: ...B CommandReference...
Страница 128: ...C CalibrationPrograms...
Страница 139: ......
Страница 141: ......