S535 Wafer Acceptance Test System Administrative Guide
Section 3: Equipment startup
S535-924-01 Rev. B / January 2019
3-3
Before starting system software
You must make sure that all of the instruments are connected with the appropriate interface cable
and a TSP-Link
®
connection between any Series 2600B, 2450, 2461-SYS, or DMM7510
instruments.
Assign GPIB or TCP/IP addresses (as appropriate) and node numbers to the hardware and
instruments.
To avoid instrument errors, make sure that all of the instruments in the system are completely
powered up and have finished self-testing before starting the system software.
The S535 system includes the following system software option:
Keithley Test Environment (KTE)
For more information about the KTE software setup procedures, refer to the
KTE S530/S535/S540
Release Notes
document (part number PA-1036).
Using Telnet
Telnet is not enabled by default on the S535 system; you must to enable it to use it.
To enable Telnet:
1.
From the command prompt (as root user), type
su
and the root password.
2.
Type
gedit /etc/xinetd.d/telnet
to open the Telnet server configuration file.
3.
In the opened file, change the line that says
disable = yes
to
disable = no
(see the following
figure).
Figure 37: Enabling Telnet