B-22
Calibration Reference
Model 2701 Service Manual
Remote error reporting
Methods to detect and determine the nature of calibration errors are discussed in the follow-
ing paragraphs. See the Model 2701 User’s Manual for details on status register operation.
Error summary
Table B-5
summarizes Model 2701 calibration errors.
Table B-5
Calibration error summary
Error number and description
+400, "10 vdc zero error"
+401, "100 vdc zero error"
+402, "10 vdc full scale error"
+403, "-10 vdc full scale error"
+404, "100 vdc full scale error"
+405, "-100 vdc full scale error"
+406, "1k 2-w zero error"
+407, "10k 2-w zero error"
+408, "100k 2-w zero error"
+409, "10M 2-w zero error "
+410, "10M 2-w full scale error"
+411, "10M 2-w open error"
+412, "1k 4-w zero error"
+413, "10k 4-w zero error"
+414, "100k 4-w zero error"
+415, "10M 4-w sense lo zero error"
+416, "1k 4-w full scale error"
+417, "10k 4-w full scale error"
+418, "100k 4-w full scale error"
+419, "1M 4-w full scale error"
+420, "10M 4-w full scale error"
+421, "10m adc zero error"
+422, "100m adc zero error"
+423, "10m adc full scale error"
+424, "100m adc full scale error"
+425, "1 adc full scale error"
+438, "Date of calibration not set"
+439, "Next date of calibration not set"
+450, "100m vac dac error"
Содержание 2701
Страница 15: ...1 PerformanceVerification...
Страница 47: ...2 Calibration...
Страница 70: ...3 RoutineMaintenance...
Страница 78: ...4 Troubleshooting...
Страница 97: ...5 Disassembly...
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 107: ......
Страница 108: ......
Страница 109: ......
Страница 110: ......
Страница 111: ...6 ReplaceableParts...
Страница 125: ......
Страница 126: ......
Страница 127: ......
Страница 128: ......
Страница 129: ......
Страница 130: ......
Страница 131: ...A Specifications...
Страница 140: ...B CalibrationReference...