Model 2701 Service Manual
Calibration
2-21
8.
After the last AC current calibration step, the instrument will prompt you to enter 3Hz
at 1V RMS and 1kHz with the following prompts:
•
Low-frequency cal
— Set the function generator to output a 1V RMS, 3Hz sine
wave. Use the left and right arrow keys and the range keys to adjust the display to
agree with the generator amplitude you measured previously, then press ENTER.
•
Frequency cal
— Set the function generator to output a 1V RMS, 1kHz sine wave.
Enter 1.000000kHz at the prompt, then press ENTER.
9.
Set the calibration dates, then save calibration to complete the process.
Remote manufacturing calibration
1.
Install the shorted Model 7797 calibration board (see “Calibration card preparation”
earlier in this section) in scanner card Slot 1 and select the rear inputs with the INPUTS
switch. Allow three minutes for thermal equilibrium.
2.
Press and hold the OPEN key while turning on the power.
3.
Enable calibration by sending the :CODE command. For example, the default
command is:
:CAL:PROT:CODE 'KI002701'
4.
Initiate calibration by sending the following command:
:CAL:PROT:INIT
5.
Calibrate step 0 with the following command:
:CAL:PROT:DC:STEP0
6.
Perform the entire remote comprehensive calibration procedure discussed earlier in this
section. (See
“Comprehensive calibration” on page 2-5
.)
7.
Connect the function generator to the Model 2701 INPUT jacks as shown in
Figure 2-5
.
Select the front input jacks with the INPUTS switch.
8.
Set the generator to output a 1V RMS, 3Hz sine wave, then send the following
command:
:CAL:PROT:AC:STEP14 <Cal_voltage>
Here <Cal_voltage> is the actual 3Hz generator signal amplitude you measured
previously.
9.
Set the generator to output a 1V RMS, 1kHz sine wave, then send the following
command:
:CAL:PROT:AC:STEP15 1E3
10.
Send the following commands to set calibration dates, save calibration, and lock out
calibration:
:CAL:PROT:DATE <year>, <month>, <day>
:CAL:PROT:NDUE <year>, <month>, <day>
:CAL:PROT:SAVE
:CAL:PROT:LOCK
Содержание 2701
Страница 15: ...1 PerformanceVerification...
Страница 47: ...2 Calibration...
Страница 70: ...3 RoutineMaintenance...
Страница 78: ...4 Troubleshooting...
Страница 97: ...5 Disassembly...
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 107: ......
Страница 108: ......
Страница 109: ......
Страница 110: ......
Страница 111: ...6 ReplaceableParts...
Страница 125: ......
Страница 126: ......
Страница 127: ......
Страница 128: ......
Страница 129: ......
Страница 130: ......
Страница 131: ...A Specifications...
Страница 140: ...B CalibrationReference...