Calibration
Model 2701 Service Manual
Measuring function generator signal amplitude
The 3Hz function generator signal amplitude must be accurately measured using the digital
multimeter listed in
Table 2-13
. Proceed as follows:
1.
Connect the function generator output to the digital multimeter INPUT jacks. (See
Figure 2-5
for typical connections.)
2.
Turn on the function generator and multimeter, allow a two-hour warm-up period
before measuring.
3.
Set the function generator to output a 1V RMS sine wave at 3Hz. Measure and record
the signal amplitude.
Front panel manufacturing calibration
1.
Install the shorted Model 7797 calibration board (see “Calibration card preparation”
earlier in this section) in scanner card Slot 1 and select the rear inputs with the INPUTS
switch. Allow three minutes for thermal equilibrium.
2.
Press and hold the OPEN key while turning on the power.
3.
Press SHIFT then TEST, then display CALIB: TEST with the up or down range key.
Press ENTER, select RUN, then enter the appropriate calibration code (default:
002701).
4.
Select ALL at the CAL:RUN prompt.
5.
Press ENTER to perform the first manufacturing calibration step.
6.
Perform the entire front panel comprehensive calibration procedure discussed earlier in
this section. (See “Comprehensive calibration” earlier in this section.)
7.
Connect the function generator to the Model 2701 front panel INPUT jacks as shown in
Figure 2-5
. Select the front input jacks with the INPUTS switch.
Figure 2-5
Function generator connections for manufacturing calibration
!
Model 2701 Ethernet Multimeter / Data Acquisition System
RANGE
F
500V
PEAK
FRONT/REAR
3A 250V
AMPS
HI
INPUT
LO
SENSE
Ω
4 WIRE
INPUTS
350V
PEAK
1000V
PEAK
AUTO
SHIFT
LOCAL
POWER
RANGE
R
EXIT
ENTER
DIGITS RATE
REL
FILTER
TRIG
EX TRIG
STORE RECALL
OPEN
DCV
DCI
MATH
O U T P U T
RATIO
ACV
ACI
Ω
2
Ω
4
FREQ
TEMP
CH AVG
CONT
PERIOD SENSOR
LIMITS
ON/OFF
DELAY
HOLD
SAVE
SETUP
CONFIG
HALT
TYPE
LSYNC
TEST
MONITOR
STEP
SCAN
OCOMP
CH-OFF
CARD
CLOSE
Integra Series
RS-232
ETHERNET
Function Generator
Model 2701
50
Ω
Coax
Cable
Function
Output
BNC-to-Dual
Banana Jack
Adapter
INPUT
HI AND LO
Note: Output voltage must be
accurately measured.
(See text.)
CAT I
Содержание 2701
Страница 15: ...1 PerformanceVerification...
Страница 47: ...2 Calibration...
Страница 70: ...3 RoutineMaintenance...
Страница 78: ...4 Troubleshooting...
Страница 97: ...5 Disassembly...
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 107: ......
Страница 108: ......
Страница 109: ......
Страница 110: ......
Страница 111: ...6 ReplaceableParts...
Страница 125: ......
Страница 126: ......
Страница 127: ......
Страница 128: ......
Страница 129: ......
Страница 130: ......
Страница 131: ...A Specifications...
Страница 140: ...B CalibrationReference...