B-20
Calibration Reference
Model 2701 Service Manual
:LOCK
(:CALibration:PROTected:CARD1:LOCK)
Purpose
To lock out Model 7700 calibration.
Format
:cal:prot:card1:lock
Parameter
None
Description
The
:LOCK
command allows you to lock out Model 7700 calibration after
completing the procedure. Thus,
:LOCK
performs the opposite of enabling
calibration with the
:CODE
command.
Example
:CAL:PROT:CARD1:LOCK
Lock out card 1 calibration.
:LOCK?
(:CALibration:PROTected:CARD1:LOCK?)
Purpose
To read Model 7700 calibration lock status.
Format
:cal:prot:card1:lock?
Response
0
Calibration locked.
1
Calibration unlocked.
Description
The
:LOCK?
query requests status from the Model 2701 on Model 7700
calibration locked/unlocked state. Calibration must be enabled sending the
:CODE
command before calibration can be performed.
Example
:CAL:PROT:CARD1:LOCK?
Request card 1 cal lock state.
:RCOunt
(:CALibration:PROTected:CARD1:RCOunt)
Purpose
To reset card calibration count to 0.
Format
:cal:prot:card1:rco
Parameter
None
Description
The
:RCOunt
command resets the card calibration count reported by
:COUNt?
to 0.
Example
:CAL:PROT:CARD1:RCO
Rest card calibration count.
Содержание 2701
Страница 15: ...1 PerformanceVerification...
Страница 47: ...2 Calibration...
Страница 70: ...3 RoutineMaintenance...
Страница 78: ...4 Troubleshooting...
Страница 97: ...5 Disassembly...
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 107: ......
Страница 108: ......
Страница 109: ......
Страница 110: ......
Страница 111: ...6 ReplaceableParts...
Страница 125: ......
Страница 126: ......
Страница 127: ......
Страница 128: ......
Страница 129: ......
Страница 130: ......
Страница 131: ...A Specifications...
Страница 140: ...B CalibrationReference...