B-12
Calibration Reference
Model 2701 Service Manual
:STEP8
(:CALibration:PROTected:DC:STEP8)
Purpose
To program the 100k
Ω
4-wire comprehensive calibration step.
Format
:cal:prot:dc:step8 <Cal_resistance>
Parameter
<Cal_resistance> = 90E3 to 110E3 [
Ω
]
Description
:STEP8
programs the 100k
Ω
4-wire resistance comprehensive calibration
step. The allowable range of the calibration resistance parameter is from
90E3 to 110E3, but 100E3 is recommended for best results.
Example
:CAL:PROT:DC:STEP8 100E3
Program 100k
Ω
step.
:STEP9
(CALibration:PROTected:DC:STEP9)
Purpose
To program the 1M
Ω
comprehensive calibration step.
Format
:cal:prot:dc:step9 <Cal_resistance>
Parameter
<Cal_resistance> = 900E3 to 1.1E6 [
Ω
]
Description
:STEP9
programs the 1M
Ω
comprehensive calibration step. The allowable
range of the calibration resistance parameter is from 900E3 to 1.1E6. Use
the 1E6 value whenever possible or the closest possible value.
Example
:CAL:PROT:DC:STEP9 1E6
Program 1M
Ω
calibration step.
:STEP10
(CALibration:PROTected:DC:STEP10)
Purpose
To program the 10mA comprehensive calibration step.
Format
:cal:prot:dc:step10 <Cal_current>
Parameter
<Cal_current> = 9E-3 to 11E-3 [A]
Description
:STEP10
programs the 10mA comprehensive calibration step. The allow-
able range of the calibration current parameter is from 9E-3 to 11E-3. Use
the 10E-3 value whenever possible for best results.
Example
:CAL:PROT:DC:STEP10 10E-3
Program 10mA step.
Содержание 2701
Страница 15: ...1 PerformanceVerification...
Страница 47: ...2 Calibration...
Страница 70: ...3 RoutineMaintenance...
Страница 78: ...4 Troubleshooting...
Страница 97: ...5 Disassembly...
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 107: ......
Страница 108: ......
Страница 109: ......
Страница 110: ......
Страница 111: ...6 ReplaceableParts...
Страница 125: ......
Страница 126: ......
Страница 127: ......
Страница 128: ......
Страница 129: ......
Страница 130: ......
Страница 131: ...A Specifications...
Страница 140: ...B CalibrationReference...