Model 2701 Service Manual
Calibration Reference
B-3
Command
Description
:CALibration
:PROTected
:DC
:STEP9 <NRf>
1M
Ω
4-wire step.
:STEP10 <NRf>
10mA DC step.
:STEP11 <NRf>
100mA DC step.
:STEP12 <NRf>
1A DC step.
:AC
AC cal steps.
:STEP1
10mV AC at 1kHz step.
:STEP2
100mV AC at 1kHz step.
:STEP3
100mV AC at 50kHz step.
:STEP4
1V AC at 1kHz step.
:STEP5
1V AC at 50kHz step.
:STEP6
10V AC at 1kHz step.
:STEP7
10V AC at 50kHz step.
:STEP8
100V AC at 1kHz step.
:STEP9
100V AC at 50kHz step.
:STEP10
700V AC at 1kHz step.
:STEP11
100mA AC at 1kHz step.
:STEP12
1A AC at 1kHz step.
:STEP13
2A AC at 1kHz step.
:STEP14
1V AC at 3Hz step
1
.
:STEP15
1V AC at 1kHz step
1
.
NOTE: Upper-case letters indicated short form of each command. For example, instead of sending
“:CALibration:PROTected:INITiate,” you can send “:CAL:PROT:INIT.”
1
DC:STEP0, AC:STEP14, and AC:STEP15 are one-time factory calibration points and are valid only in manufacturing calibration
mode.
Table B-1 (Continued)
Remote calibration command summary
Содержание 2701
Страница 15: ...1 PerformanceVerification...
Страница 47: ...2 Calibration...
Страница 70: ...3 RoutineMaintenance...
Страница 78: ...4 Troubleshooting...
Страница 97: ...5 Disassembly...
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 107: ......
Страница 108: ......
Страница 109: ......
Страница 110: ......
Страница 111: ...6 ReplaceableParts...
Страница 125: ......
Страница 126: ......
Страница 127: ......
Страница 128: ......
Страница 129: ......
Страница 130: ......
Страница 131: ...A Specifications...
Страница 140: ...B CalibrationReference...