2-16
Calibration
Model 2701 Service Manual
DC current calibration
After the 1M
Ω
resistance point has been calibrated, follow these steps for DC current
calibration:
1.
Connect the calibrator to the AMPS and INPUT LO terminals of the Model 2701 as
shown in
Figure 2-3
.
2.
Perform the calibration steps listed in
Table 2-10
. For each step:
•
Set the calibrator to the indicated current and make sure the unit is in operate. (Use
the recommended current if possible.)
•
Send the indicated programming command. (Change the current parameter if you
are using a different calibration current.)
•
Wait until the Model 2701 completes each step before continuing.
NOTE
If you are performing DC-only calibration, proceed to
“Programming calibration
dates” on page 2-18
Table 2-10
DC current calibration programming steps
Calibration
step
Calibrator
current
Calibration command*
Parameter range
10mA
100mA
1A
10.00000mA
100.00000mA
1.000000A
:CAL:PROT:DC:STEP10 10E-3
:CAL:PROT:DC:STEP11 100E-3
:CAL:PROT:DC:STEP12 1
9E-3 to 11E-3
90E-3 to 110E-3
0.9 to 1.1
*Change parameter if using different current.
Содержание 2701
Страница 15: ...1 PerformanceVerification...
Страница 47: ...2 Calibration...
Страница 70: ...3 RoutineMaintenance...
Страница 78: ...4 Troubleshooting...
Страница 97: ...5 Disassembly...
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 107: ......
Страница 108: ......
Страница 109: ......
Страница 110: ......
Страница 111: ...6 ReplaceableParts...
Страница 125: ......
Страница 126: ......
Страница 127: ......
Страница 128: ......
Страница 129: ......
Страница 130: ......
Страница 131: ...A Specifications...
Страница 140: ...B CalibrationReference...