Section 7: SCPI command reference
Series 2280 Precision DC Power Supply Reference Manual
7-2
077085503 / March 2019
Although the commands in this section are shown in uppercase, common commands are not case
sensitive (you can use either uppercase or lowercase).
*CLS
This command clears the event registers and error queues.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
*CLS
Details
Use
*CLS
to clear the event registers and queues. It performs the following actions:
Clear the Standard Event Status Register and Status Byte Register.
Clear the event register of Questionable Event, Questionable Instrument Event, and Questionable
Instrument Summary Event.
Clear the event register of Operation Event, Operation Instrument Event, and Operation Instrument
Summary Event.
Clear the event register of Measurement Event, Measurement Instrument Event, and Measurement
Instrument Summary Event.
Clear the error queue.
To reset all of the bits of the Standard Event Enable Register, send the command:
*ESE 0
Also see
(on page 7-3)
(on page 7-94)