Series 2280 Precision DC Power Supply Reference Manual
Section 3: Functions and features
077085503 / March 2019
3-11
Using the front panel to set fill mode:
1. Press the
MENU
key.
2. Under Measure, select
Data Buffers
.
3. Select the button next to
Mode
. The Mode screen is displayed.
4. Select the buffer Mode. For example, press the button next to
Always
.
5. Press the
HOME
key to return to the home screen.
To set the reading buffer mode to overwrite old readings when the buffer is full, send the following
command:
:TRACe:FEED:CONTrol ALWays
To set the reading buffer mode to start but not overwrite old readings, send the following command:
:TRACe:FEED:CONTrol NEXT
To set the reading buffer mode so that no readings are saved in the buffer, send the following
command:
:TRACe:FEED:CONTrol NEVer
To set the reading buffer to hold 300 readings, send the following command:
:TRACe:POINts 300