Section 7: SCPI command reference
Series 2280 Precision DC Power Supply Reference Manual
7-96
077085503 / March 2019
Example
:STAT:MEAS:ENAB #b0010000000000000
Sets the ISUM bit of the measurement
status enable register using a binary
value.
You could also send the decimal value
8192
or the hexadecimal value:
#h2000
.
Also see
(on page D-17)
:STATus:MEASurement:INSTrument[:EVENt]?
This command reads and clears the Measurement Instrument Event Register of the status model.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:STATus:MEASurement:INSTrument[:EVENt]?
Details
This attribute reads the Measurement Instrument Event Register of the status model and then clears
the register.
The instrument returns a decimal value that corresponds to the binary-weighted sum of all bits set in
the register. For details, see
Measurement Instrument Event Register
(on page D-18).
Also see
None