Series 2280 Precision DC Power Supply Reference Manual
Section 2: General operation
077085503 / March 2019
2-65
Using USB
To communicate from a computer to the instrument you need a USB cable with a USB Type B
connector end and a USB type A connector end. You need a separate USB cable for each instrument
you plan to connect to the computer at the same time using the USB interface.
To connect an instrument to a computer using USB:
1. Connect the Type A end of the cable to the computer.
2. Connect the Type B end of the cable to the instrument.
3. Turn on the instrument power. When the computer detects the new USB connection, the Found
New Hardware Wizard starts.
4. If the "Can Windows connect to Windows Update to search for software?" dialog box opens,
select
No
, and then select
Next
.
5. On the "USB Test and Measurement device" dialog box, select
Next
, and then select
Finish
.
Communicate with the instrument
For the instrument to communicate with the USB device, you must use NI-VISA
TM
. VISA requires a
resource string in the following format to connect to the correct USB instrument:
USB0::0x05E6::0x2280::[serial number]::INSTR
Where:
0x05e6
= The Keithley Instruments vendor ID
0x2280
= The instrument model number
[serial number]
= The serial number of the instrument (the serial number is also on the rear
panel)
INSTR
= Use the USBTMC protocol
To determine these parameters, you can run the Keithley Configuration Panel, which automatically
detects all instruments connected to the computer.
If you installed the Keithley I/O Layer, you can access the Keithley Configuration Panel through the
Microsoft
®
Windows
®
Start menu.