Section 2: General operation
Series 2280 Precision DC Power Supply Reference Manual
2-48
077085503 / March 2019
The address is saved in nonvolatile memory, so it does not change when a reset is done or when the
instrument power is turned off and then turned on again.
To set the GPIB address from the front panel:
1. Press the
MENU
key.
2. Under System, select
Communication
. The SYSTEM COMMUNICATION window opens.
3. Select the
GPIB
interface.
4. Next to Address, select the number. The Set GPIB Address dialog box is displayed.
5. Enter the address.
6. Select
OK
.
If you are using a Series 2280 with no front panel, you can set the GPIB address with the SCPI
command
:SYSTem:GPIB:ADDRess
.
LAN communications
The Series 2280 comes with one standard LAN crossover cable for LAN communication. However,
you can use any standard LAN crossover cable (RJ-45, male-to-male) or straight-through cable to
connect your equipment. The instrument automatically senses which cable you have connected.
The following figure shows the location of the LAN connection on the rear panel of the instrument.
Connect the LAN cable between this connection and the LAN connector on the computer.