Series 2280 Precision DC Power Supply Reference Manual
Section 7: SCPI command reference
077085503 / March 2019
7-99
:STATus:MEASurement:INSTrument:ISUMmary:ENABle
This command sets or reads the contents of the Measurement Instrument Summary Enable Register of the status
model.
Type
Affected by
Where saved
Default value
Command and query
Power cycle
Not applicable
0
Usage
:STATus:MEASurement:INSTrument:ISUMmary:ENABle <n>
:STATus:MEASurement:INSTrument:ISUMmary:ENABle?
[n]
The status of the Measurement Instrument Summary Enable Register.
Details
This command sets or reads the contents of the enable register of the Measurement Instrument
Summary Event Register (MISER).
When one of these bits is set and the corresponding bit in the MISER set, bit 1 in the Measurement
Instrument Event Register is set.
When sending binary, preface
<n>
with
#b
. When sending hexadecimal, preface
<n>
with
#h
. No
preface is needed when sending decimal values.
Example
:STAT:MEAS:INST:ISUM:ENAB #b0000100000000001
Sets the output and reading overflow bits
of the Measurement Instrument
Summary Enable register using a binary
value.
You could also send the decimal value
2049
or the hexadecimal value:
#h801
.
Also see
Measurement Instrument Summary Event Register
(on page D-19)