Ratio and Delta
5-5
NOTE
The previous calculation shows Filter enabled. If Filter is not used, remove the “Filt”
component from the calculation.
When Ratio is enabled, the state (on or off) of the REL annunciator depends on which
measurement function was last selected. If on DCV1 when Ratio is enabled, the state of the REL
annunciator (on or off) will indicate the state (enabled or disabled) of Rel for DCV1. If on DCV2
when Ratio is enabled, the state if the REL annunciator will indicate the state of Rel for DCV2.
The REL key is operational while in Ratio. Pressing REL will either disable Rel for both
channels or enable Rel for both channels (REL annunciator turns on). When Rel is enabled, the
instrument acquires the input signal from each of the two channels as Rel values. Each Rel value
is then applied to the respective channel. Keep in mind that the Rel operations are performed on
the input channels, not on the result of Ratio.
Ranging considerations
As explained in
, a separate range setting (fixed or AUTO) can be used for each
voltage channel. When Ratio is enabled, the range setting for each channel is retained. For
example, Channel 1 could be set for autoranging, and Channel 2 could be fixed on the 10V
range.
Range control
— The manual range keys can only control one of the two channels. If the
instrument is on DCV1, TEMP1, or TEMP2 when Ratio is enabled, the manual range keys will
control Channel 1 (DCV1). The manual range keys will have no effect on Channel 2 (DCV2). If
on DCV2 (Channel 2) when Ratio is enabled, range control will apply to Channel 2 (DCV2).
The manual range keys will have no effect on Channel 1 (DCV1).
NOTES
• When Ratio is selected, the range control channel will be displayed while the
“CH1 / CH2” message is being displayed.
• When a range key is pressed, the channel number annunciator for the range
controlled channel remains on. The other channel annunciator turns off for a
brief moment.
• The state (on or off) of the AUTO range annunciator indicates the state
(enabled or disabled) of autorange for the voltage channel that is under range
control.
• With Ratio already enabled, pressing the AUTO range key will either disable
autorange for both channels or enable autorange for both channels (AUTO
annunciator turns on).
Содержание 2182
Страница 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Страница 18: ......
Страница 22: ......
Страница 23: ...1 Getting Started Getting Started...
Страница 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Страница 68: ...2 28 Voltage and Temperature Measurements...
Страница 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Страница 82: ...3 14 Range Digits Rate and Filter...
Страница 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Страница 91: ...5 RatioandDelta Ratioand Delta...
Страница 117: ...6 Buffer Buffer...
Страница 123: ...7 Triggering Triggering...
Страница 140: ...7 18 Triggering...
Страница 141: ...8 Limits Limits...
Страница 149: ...9 SteppingandScanning Steppingand Scanning...
Страница 168: ...9 20 Stepping and Scanning...
Страница 169: ...10 AnalogOutput Analog Output...
Страница 175: ...11 RemoteOperation Remote Operation...
Страница 205: ...12 CommonCommands Common Commands...
Страница 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Страница 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Страница 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Страница 260: ...15 22 Additional SCPI Commands...
Страница 261: ...A Specifications Specifications...
Страница 263: ...B Statusand ErrorMessages StatusandError Messages...
Страница 268: ...B 6 Status and Error Messages...
Страница 269: ...C Measurement Considerations Measurement Consider ations...
Страница 278: ...C 10 Measurement Considerations...
Страница 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Страница 284: ...D 6 Model 182 Emulation Commands...
Страница 285: ...E Example Programs ExamplePro grams...
Страница 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Страница 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Страница 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Страница 311: ...H Measurement Queries Measurement Queries...
Страница 316: ...H 6 Measurement Queries...
Страница 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...