Remote Operation
11-27
Message exchange protocol
Two rules summarize the message exchange protocol:
Rule 1.
Always tell the Model 2182 what to send to the computer.
The following two steps must always be performed to send information from the
instrument to the computer:
1.
Send the appropriate query command(s) in a program message.
2.
Address the Model 2182 to talk.
Rule 2.
The complete response message must be received by the computer before another
program message can be sent to the Model 2182.
RS-232 interface reference
Sending and receiving data
The RS-232 interface transfers data using eight data bits, one stop bit, and no parity. Make
sure the controller you connect to the Model 2182 also uses these settings.
You can break data transmissions by sending a ^C or ^X character string to the controller.
This clears any pending operation and discards any pending output.
Baud rate, flow control and terminator
NOTE
The procedure to select and configure the RS-232 interface is provided in “
” located at the beginning of this section.
Baud rate
The baud rate is the rate at which the Model 2182 and the programming terminal
communicate. Choose one of the following available rates:
•
19.2k
•
9600
•
4800
•
2400
•
1200
•
600
•
300
The factory selected baud rate is 9600.
Make sure that the programming terminal that you are connecting to the Model 2182 can
support the baud rate you selected. Both the Model 2182 and the other device must be configured
for the same baud rate. To select a baud rate, follow these steps:
Содержание 2182
Страница 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Страница 18: ......
Страница 22: ......
Страница 23: ...1 Getting Started Getting Started...
Страница 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Страница 68: ...2 28 Voltage and Temperature Measurements...
Страница 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Страница 82: ...3 14 Range Digits Rate and Filter...
Страница 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Страница 91: ...5 RatioandDelta Ratioand Delta...
Страница 117: ...6 Buffer Buffer...
Страница 123: ...7 Triggering Triggering...
Страница 140: ...7 18 Triggering...
Страница 141: ...8 Limits Limits...
Страница 149: ...9 SteppingandScanning Steppingand Scanning...
Страница 168: ...9 20 Stepping and Scanning...
Страница 169: ...10 AnalogOutput Analog Output...
Страница 175: ...11 RemoteOperation Remote Operation...
Страница 205: ...12 CommonCommands Common Commands...
Страница 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Страница 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Страница 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Страница 260: ...15 22 Additional SCPI Commands...
Страница 261: ...A Specifications Specifications...
Страница 263: ...B Statusand ErrorMessages StatusandError Messages...
Страница 268: ...B 6 Status and Error Messages...
Страница 269: ...C Measurement Considerations Measurement Consider ations...
Страница 278: ...C 10 Measurement Considerations...
Страница 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Страница 284: ...D 6 Model 182 Emulation Commands...
Страница 285: ...E Example Programs ExamplePro grams...
Страница 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Страница 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Страница 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Страница 311: ...H Measurement Queries Measurement Queries...
Страница 316: ...H 6 Measurement Queries...
Страница 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...