Range, Digits, Rate, and Filter
3-3
Range
The selected range affects both accuracy of the voltage measurement as well as the maximum
voltage that can be measured. The DCV1 function has five measurement ranges; 10mV, 100mV,
1V, 10V, and 100V. The DCV2 function has three measurement ranges; 100mV, 1V, and 10V.
The range setting (fixed or AUTO) is remembered by each voltage function.
NOTE
The available voltage ranges for Ratio (V1/V2) depend on which channel is presently
selected when Ratio is enabled. If Channel 1 is presently selected, DCV1 ranges will
be available when Ratio is enabled. If Channel 2 is presently selected, DCV2 ranges
will be available when Ratio is enabled. Complete information on ranging for Ratio
is provided in
There is no range selection for temperature (TEMP1 and TEMP2) measurements.
Temperature measurements are performed on a single, fixed range. The DIGITS key sets reading
resolution.
Maximum readings
The full scale readings for every voltage range are 20% over range. For example, on the 10V
range, the maximum input voltage is ±12V.
Depending on which type of thermocouple is being used, the maximum temperature readings
range from –200˚C to 1820˚C. The Specifications (
) list the reading range for each
thermocouple type.
Input values that exceed the maximum readings cause the overflow message (“OVRFLW”)
to be displayed.
Manual ranging
To select a range, press the RANGE
or
key. The instrument changes one range per
key-press. The selected range is displayed for one second. Note that the manual range keys have
no effect on temperature (TEMP1 and TEMP2).
If the instrument displays the “OVRFLW” message on a particular range, select a higher
range until an on-range reading is displayed. Use the lowest range possible without causing an
overflow to ensure best accuracy and resolution.
Содержание 2182
Страница 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Страница 18: ......
Страница 22: ......
Страница 23: ...1 Getting Started Getting Started...
Страница 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Страница 68: ...2 28 Voltage and Temperature Measurements...
Страница 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Страница 82: ...3 14 Range Digits Rate and Filter...
Страница 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Страница 91: ...5 RatioandDelta Ratioand Delta...
Страница 117: ...6 Buffer Buffer...
Страница 123: ...7 Triggering Triggering...
Страница 140: ...7 18 Triggering...
Страница 141: ...8 Limits Limits...
Страница 149: ...9 SteppingandScanning Steppingand Scanning...
Страница 168: ...9 20 Stepping and Scanning...
Страница 169: ...10 AnalogOutput Analog Output...
Страница 175: ...11 RemoteOperation Remote Operation...
Страница 205: ...12 CommonCommands Common Commands...
Страница 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Страница 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Страница 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Страница 260: ...15 22 Additional SCPI Commands...
Страница 261: ...A Specifications Specifications...
Страница 263: ...B Statusand ErrorMessages StatusandError Messages...
Страница 268: ...B 6 Status and Error Messages...
Страница 269: ...C Measurement Considerations Measurement Consider ations...
Страница 278: ...C 10 Measurement Considerations...
Страница 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Страница 284: ...D 6 Model 182 Emulation Commands...
Страница 285: ...E Example Programs ExamplePro grams...
Страница 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Страница 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Страница 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Страница 311: ...H Measurement Queries Measurement Queries...
Страница 316: ...H 6 Measurement Queries...
Страница 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...