5-22
Ratio and Delta
Superconductor Application #2 - fixed magnetic field
Another typical test on a superconductor sample (DUT) is to source an increasing-amplitude
current (I) through the DUT, while maintaining the magnetic field (H) at a fixed level. The I-V
curve in
shows that the measured voltage across the DUT remains at ~0V for low
currents (I). This is the flat portion of the curve where the DUT remains at 0
Ω
. At some point,
the voltage drop across the DUT will start increasing as current through the DUT increases. The
actual resistance of the DUT can be calculated at any current source point using Ohm’s law:
R = V/I
where: R is actual resistance of the DUT.
V is the measured voltage across the DUT.
I is the known current that flows through the DUT.
Figure 5-9
I-V Curve (Fixed H)
Such a test system is shown in
. A Keithley SourceMeter (Model 2400, 2410, or
2420) is used to source current through the DUT and two Model 2182s are used to provide
simultaneous voltage measurements.
Model 2182 #1 measures the voltage across a precision reference resistor (R
REF
) and stores
the readings in its buffer. These stored readings allow you to reference current amplitudes to the
voltage measurements of the DUT. Model 2182 #2 measures voltage across the DUT and stores
the readings in its buffer.
For example, assume you want to measure DUT voltage at current sweep values of 10µA,
20µA, and 50µA. When the sweep is started (10µA output), Model 2182 #1 measures 1mV
(10µA
×
100
Ω
= 1mV) and stores the reading in its buffer at location 1. At the same time,
Model 2182 #2 measures the DUT and stores that reading in its buffer at location 1. At the next
sweep point (20µA), Model 2182 #1 measures 2mV and stores the reading in its buffer at
location 2, and Model 2182 #2 measures the DUT and stores the reading in its buffer at location
2. At the last sweep point (50µA), Model 2182 #1 measures 5mV and stores the reading in its
buffer at location 3, and Model 2182 #2 measures the DUT and stores the reading in its buffer at
location 3.
Measure
V
Fixed H
Source I
Содержание 2182
Страница 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Страница 18: ......
Страница 22: ......
Страница 23: ...1 Getting Started Getting Started...
Страница 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Страница 68: ...2 28 Voltage and Temperature Measurements...
Страница 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Страница 82: ...3 14 Range Digits Rate and Filter...
Страница 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Страница 91: ...5 RatioandDelta Ratioand Delta...
Страница 117: ...6 Buffer Buffer...
Страница 123: ...7 Triggering Triggering...
Страница 140: ...7 18 Triggering...
Страница 141: ...8 Limits Limits...
Страница 149: ...9 SteppingandScanning Steppingand Scanning...
Страница 168: ...9 20 Stepping and Scanning...
Страница 169: ...10 AnalogOutput Analog Output...
Страница 175: ...11 RemoteOperation Remote Operation...
Страница 205: ...12 CommonCommands Common Commands...
Страница 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Страница 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Страница 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Страница 260: ...15 22 Additional SCPI Commands...
Страница 261: ...A Specifications Specifications...
Страница 263: ...B Statusand ErrorMessages StatusandError Messages...
Страница 268: ...B 6 Status and Error Messages...
Страница 269: ...C Measurement Considerations Measurement Consider ations...
Страница 278: ...C 10 Measurement Considerations...
Страница 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Страница 284: ...D 6 Model 182 Emulation Commands...
Страница 285: ...E Example Programs ExamplePro grams...
Страница 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Страница 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Страница 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Страница 311: ...H Measurement Queries Measurement Queries...
Страница 316: ...H 6 Measurement Queries...
Страница 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...