Delta, Pulse Delta, and Differential Conductance
I-15
Figure I-7
Differential Conductance measurement process
10µA
[(A-B)/2] + [(C-B)/2]
2
Start
Stop
Step
dV #1
(-1)
0
=
·
Step
Step
Step
Step
Step
dV Calc
#1
dV Calc
#4
Start = 0µA
Step = 10µA
Stop = 50µA
Delta (dI) = 20µA
[(B-C)/2] + [(D-C)/2]
2
dV #2
(-1)
1
=
·
[(C-D)/2] + [(E-D)/2]
2
dV #3
(-1)
2
=
·
[(D-E)/2] + [(F-E)/2]
2
dV #4
(-1)
3
=
·
[(E-F)/2] + [(G-F)/2]
2
dV #5
(-1)
4
=
·
[(F-G)/2] + [(H-G)/2]
2
dV #6
(-1)
5
=
·
= Sweep with 20µA Delta (dI)
= Sweep with 0µA Delta (dI)
time
0µA
20µA
30µA
40µA
50µA
60µA
70µA
-10µA
Step
Delta
Delta
Delta
Delta
Delta
Delta
Delta
Delta
dV Calc
#3
dV Calc
#5
dV Calc
#2
dV Calc
#6
-20µA
dV Calc #2 B to D
dV Calc #3 C to E
dV Calc #4 D to F
dV Calc #5 E to G
dV Calc #6 F to H
dV Calc #1 A to C
dV Calculations
The following equations are used by the 622x to calculate differential voltage (dV).
To calculate dV, points A through H are 2182/2182A voltage measurements (A/D readings).
A/D Rdg
A
A/D Rdg C
A/D Rdg E
A/D Rdg H
A/D Rdg F
A/D Rdg D
A/D Rdg = 2182/2182A voltage measurement conversion.
dG and dR Calculations
With dI known (dI = Delta) and dV calculated, the 622x can then calculate
differential conductance (dG) or differential resistance (dR).
With G units selected, readings are calculated as follows: dG = dI/dV.
With R units selected, readings are calculated as follows: dR = dV/dI.
dV Calc = Calculate differential voltage (dV) using
last three A/D Rdgs.
A/D Rdg G
A/D Rdg B
Содержание 2182
Страница 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Страница 18: ......
Страница 22: ......
Страница 23: ...1 Getting Started Getting Started...
Страница 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Страница 68: ...2 28 Voltage and Temperature Measurements...
Страница 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Страница 82: ...3 14 Range Digits Rate and Filter...
Страница 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Страница 91: ...5 RatioandDelta Ratioand Delta...
Страница 117: ...6 Buffer Buffer...
Страница 123: ...7 Triggering Triggering...
Страница 140: ...7 18 Triggering...
Страница 141: ...8 Limits Limits...
Страница 149: ...9 SteppingandScanning Steppingand Scanning...
Страница 168: ...9 20 Stepping and Scanning...
Страница 169: ...10 AnalogOutput Analog Output...
Страница 175: ...11 RemoteOperation Remote Operation...
Страница 205: ...12 CommonCommands Common Commands...
Страница 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Страница 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Страница 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Страница 260: ...15 22 Additional SCPI Commands...
Страница 261: ...A Specifications Specifications...
Страница 263: ...B Statusand ErrorMessages StatusandError Messages...
Страница 268: ...B 6 Status and Error Messages...
Страница 269: ...C Measurement Considerations Measurement Consider ations...
Страница 278: ...C 10 Measurement Considerations...
Страница 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Страница 284: ...D 6 Model 182 Emulation Commands...
Страница 285: ...E Example Programs ExamplePro grams...
Страница 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Страница 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Страница 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Страница 311: ...H Measurement Queries Measurement Queries...
Страница 316: ...H 6 Measurement Queries...
Страница 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...