Additional SCPI Commands
15-3
DISPlay subsystem
The commands in this subsystem are used to control the display of the Model 2182 and are
summarized in
:ENABle <b>
:DISPlay:ENABle <b>
Control display circuitry
Parameters
<b> =
0 or OFF
Disable display circuitry
1 or ON
Enable display circuitry
Description
This command is used to enable and disable the front panel display circuitry. When disabled,
the instrument operates at a higher speed. While disabled, the display is frozen.
All front panel controls (except LOCAL) are disabled. Normal display operation can be
resumed by using the :ENABle command to enable the display or by putting the Model 2182
into local mode (press LOCAL).
:TEXT commands
:DATA <a>
:DISPlay[:WINDow[1]]:TEXT:DATA <a>
Define message for display
Parameter
<a> =
ASCII characters for the message (maximum of 12 characters). The characters
must be enclosed in either double quotes (“ ”) or single quotes (‘ ’).
Description
These commands define the text message for display. A message can be as long as 12
characters. A space counts as a character. Excess message characters results in an error.
:STATe <b>
:DISPlay[WINDow[1]]:TEXT:STATe <b>
Control (on/off) message
Parameters
<b> =
0 or OFF
Disable text message
1 or ON
Enable text message
Description
This command enables and disables the text message mode. When enabled, a defined
message is displayed. When disabled, the message is removed from the display.
A user defined text message remains displayed only as long as the instrument is in remote.
Taking the instrument out of remote (by pressing the LOCAL key or sending GTL), cancels the
message and disables the text message mode.
Содержание 2182
Страница 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Страница 18: ......
Страница 22: ......
Страница 23: ...1 Getting Started Getting Started...
Страница 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Страница 68: ...2 28 Voltage and Temperature Measurements...
Страница 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Страница 82: ...3 14 Range Digits Rate and Filter...
Страница 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Страница 91: ...5 RatioandDelta Ratioand Delta...
Страница 117: ...6 Buffer Buffer...
Страница 123: ...7 Triggering Triggering...
Страница 140: ...7 18 Triggering...
Страница 141: ...8 Limits Limits...
Страница 149: ...9 SteppingandScanning Steppingand Scanning...
Страница 168: ...9 20 Stepping and Scanning...
Страница 169: ...10 AnalogOutput Analog Output...
Страница 175: ...11 RemoteOperation Remote Operation...
Страница 205: ...12 CommonCommands Common Commands...
Страница 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Страница 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Страница 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Страница 260: ...15 22 Additional SCPI Commands...
Страница 261: ...A Specifications Specifications...
Страница 263: ...B Statusand ErrorMessages StatusandError Messages...
Страница 268: ...B 6 Status and Error Messages...
Страница 269: ...C Measurement Considerations Measurement Consider ations...
Страница 278: ...C 10 Measurement Considerations...
Страница 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Страница 284: ...D 6 Model 182 Emulation Commands...
Страница 285: ...E Example Programs ExamplePro grams...
Страница 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Страница 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Страница 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Страница 311: ...H Measurement Queries Measurement Queries...
Страница 316: ...H 6 Measurement Queries...
Страница 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...