SCPI Signal Oriented Measurement Commands
13-3
•
Buffer operation is disabled. A storage operation presently in process will be aborted.
•
Autozero is set to the *RST default value.
•
All operations associated with stepping or scanning are disabled.
This command is automatically asserted when the :MEASure? command is sent.
Program fragment
CALL SEND (7, “:conf:volt”, status%)
‘ Perform CONFigure operations.
CALL SEND (7, “:trig:del 0.5”, status%)
‘ Set delay for 0.5sec.
CALL SEND (7, “:samp:coun 10”, status%)
‘ Set sample count to 10.
CALL SEND (7, “:read?”, status%)
‘ Trigger and request readings.
reading$ = SPACE$(80)
CALL ENTER(reading$, length%, 7, status%) ‘ Address 2182 to talk.
PRINT reading$
‘ Display the 10 readings on the
‘ CRT.
:FETCh?
Description
This command requests the latest post-processed reading. After sending this command and
addressing the Model 2182 to talk, the reading is sent to the computer. This command does not
affect the instrument setup.
This command does not trigger a measurement. The command simply requests the last
available reading. Note that this command can repeatedly return the same reading. Until there is
a new reading, this command continues to return the old reading. If your application requires a
“fresh” reading, use the :DATA:FRESh? command (see the SENSe Subsystem command).
This command is automatically asserted when the :READ? or :MEASure? command is sent.
:READ?
Description
This command is typically used with the instrument in the “one-shot” measurement mode to
trigger and acquire a specified number of readings. The :SAMPle:COUNt command is used to
specify the number of readings (see Trigger Subsystem). Note that the readings are stored in the
buffer.
When this command is sent, the following commands execute in the order they are presented:
:ABORt
:INITiate
:FETCh?
When :ABORt is executed, the instrument goes into the idle state if continuous initiation is
disabled. If continuous initiation is enabled, the operation re-starts at the beginning of the
Trigger Model.
Содержание 2182
Страница 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Страница 18: ......
Страница 22: ......
Страница 23: ...1 Getting Started Getting Started...
Страница 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Страница 68: ...2 28 Voltage and Temperature Measurements...
Страница 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Страница 82: ...3 14 Range Digits Rate and Filter...
Страница 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Страница 91: ...5 RatioandDelta Ratioand Delta...
Страница 117: ...6 Buffer Buffer...
Страница 123: ...7 Triggering Triggering...
Страница 140: ...7 18 Triggering...
Страница 141: ...8 Limits Limits...
Страница 149: ...9 SteppingandScanning Steppingand Scanning...
Страница 168: ...9 20 Stepping and Scanning...
Страница 169: ...10 AnalogOutput Analog Output...
Страница 175: ...11 RemoteOperation Remote Operation...
Страница 205: ...12 CommonCommands Common Commands...
Страница 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Страница 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Страница 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Страница 260: ...15 22 Additional SCPI Commands...
Страница 261: ...A Specifications Specifications...
Страница 263: ...B Statusand ErrorMessages StatusandError Messages...
Страница 268: ...B 6 Status and Error Messages...
Страница 269: ...C Measurement Considerations Measurement Consider ations...
Страница 278: ...C 10 Measurement Considerations...
Страница 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Страница 284: ...D 6 Model 182 Emulation Commands...
Страница 285: ...E Example Programs ExamplePro grams...
Страница 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Страница 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Страница 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Страница 311: ...H Measurement Queries Measurement Queries...
Страница 316: ...H 6 Measurement Queries...
Страница 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...