NOTES
1. The fail message (“0”) for a limit test indicates that the reading is outside the
specified limits.
2. With auto clear enabled, the fail message (“0”) is cleared when the instrument
goes back into the idle state. If programmed not to go back into idle, you can
manually clear the fail condition by sending the CLEar[:IMMediate] command.
With auto clear disabled, the fail condition will have to be cleared manually.
3. When not in a continuous measurement mode (waiting for a trigger), you can
change the limits and re-test the last reading. After changing the limits, send
CALCulate3:IMMediate to re-perform the limit tests on the last reading. Note that
sending IMMediate does not initiate (trigger) a reading. It simply repeats the limit
tests on the last reading.
Programming example
The following program fragment performs limit tests on a voltage input to Channel 1:
‘Configure 2182 for one-shot DCV1 measurements:
CALL SEND(7, “*rst”,status%)
‘Put 2182 in a one-shot mode.
‘ Configure limit tests:
CALL SEND(7,“:calc3:lim:upp 0.1”,status%)
‘Set HI1 limit to 0.1.
CALL SEND(7,“:calc3:lim:low -0.1”,status%)
‘Set LO1 limit to -0.1.
CALL SEND(7,“:calc3:lim2:upp 1”,status%)
‘Set HI2 limit to 1.
CALL SEND(7,“:calc3:lim2:low -1”,status%)
‘Set LO2 limit to -1.
CALL SEND(7,“:calc3:lim:stat on”,status%)
‘Enable Limit 1 test.
CALL SEND(7,“:calc3:lim2:stat on”,status%)
‘Enable Limit 2 test.
‘Trigger one reading and display it:
CALL SEND(7,“:read?”,status%)
‘Trigger and request a
‘reading.
reading$ = SPACE$(80)
CALL ENTER(reading$, length%, 7,status%)
‘Address 2182 to talk.
PRINT reading$
‘Display reading on CRT.
‘Check result of Limit 1 test:
CALL SEND(7,“:calc3:lim:fail?”,status%)
‘Request result of
‘Limit 1 test.
reading$ = SPACE$(80)
CALL ENTER(reading$, length%, 7, status%)
‘Address 2182 to talk.
PRINT reading$
‘Display result of
‘Limit 1 test.
‘Check result of Limit 2 test:
CALL SEND (7, “:calc3:lim2:fail?”, status%)
‘Request result of
‘Limit 2 test.
reading$ = SPACE$(80)
CALL ENTER(reading$, length%, 7,status%)
‘Address 2182 to talk.
PRINT reading$
‘Display result of
‘Limit 2 test.
8-6
Limits
Содержание 2182
Страница 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Страница 18: ......
Страница 22: ......
Страница 23: ...1 Getting Started Getting Started...
Страница 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Страница 68: ...2 28 Voltage and Temperature Measurements...
Страница 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Страница 82: ...3 14 Range Digits Rate and Filter...
Страница 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Страница 91: ...5 RatioandDelta Ratioand Delta...
Страница 117: ...6 Buffer Buffer...
Страница 123: ...7 Triggering Triggering...
Страница 140: ...7 18 Triggering...
Страница 141: ...8 Limits Limits...
Страница 149: ...9 SteppingandScanning Steppingand Scanning...
Страница 168: ...9 20 Stepping and Scanning...
Страница 169: ...10 AnalogOutput Analog Output...
Страница 175: ...11 RemoteOperation Remote Operation...
Страница 205: ...12 CommonCommands Common Commands...
Страница 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Страница 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Страница 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Страница 260: ...15 22 Additional SCPI Commands...
Страница 261: ...A Specifications Specifications...
Страница 263: ...B Statusand ErrorMessages StatusandError Messages...
Страница 268: ...B 6 Status and Error Messages...
Страница 269: ...C Measurement Considerations Measurement Consider ations...
Страница 278: ...C 10 Measurement Considerations...
Страница 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Страница 284: ...D 6 Model 182 Emulation Commands...
Страница 285: ...E Example Programs ExamplePro grams...
Страница 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Страница 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Страница 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Страница 311: ...H Measurement Queries Measurement Queries...
Страница 316: ...H 6 Measurement Queries...
Страница 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...