15-14
Additional SCPI Commands
:PRESet command
:PRESet
:STATus:PRESet
Return registers to default conditions
Description
When this command is sent, all bits of the following registers are cleared to zero (0):
•
Questionable Event Enable Register
•
Measurement Event Enable Register
•
Operation Event Enable Register
NOTE
Registers not included in the above list are not affected by this command.
:QUEue commands
[:NEXT]?
:STATus:QUEue[:NEXT]?
Read Error Queue
Description
As error and status messages occur, they are placed into the Error Queue. This query
command is used to read those messages.
The Error Queue is a first-in, first-out (FIFO) register. Each time you read the queue, the
“oldest” message is read and that message is then removed from the queue. The queue will hold
up to ten messages. If the queue becomes full, the “350, Queue Overflow” message will occupy
the last memory location in the register. On power-up, the Error Queue is empty. When the Error
Queue is empty, the “0, No error” message is placed in the Error Queue.
The messages in the queue are preceded by a number. Negative (–) numbers are used for SCPI
defined messages, and positive (+) numbers are used for Keithley defined messages. The
messages are listed in
After this command is sent and the Model 2182 is addressed to talk, the “oldest” message in
the queue is sent to the computer.
NOTE
The :STATus:QUEue[:NEXT]? query command performs the same function as the
:SYSTem:ERRor? query command (see System subsystem).
:CLEar
:STATus:QUEue:CLEar
Clear Error Queue
Description
This action command is used to clear the Error Queue of messages.
Содержание 2182
Страница 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Страница 18: ......
Страница 22: ......
Страница 23: ...1 Getting Started Getting Started...
Страница 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Страница 68: ...2 28 Voltage and Temperature Measurements...
Страница 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Страница 82: ...3 14 Range Digits Rate and Filter...
Страница 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Страница 91: ...5 RatioandDelta Ratioand Delta...
Страница 117: ...6 Buffer Buffer...
Страница 123: ...7 Triggering Triggering...
Страница 140: ...7 18 Triggering...
Страница 141: ...8 Limits Limits...
Страница 149: ...9 SteppingandScanning Steppingand Scanning...
Страница 168: ...9 20 Stepping and Scanning...
Страница 169: ...10 AnalogOutput Analog Output...
Страница 175: ...11 RemoteOperation Remote Operation...
Страница 205: ...12 CommonCommands Common Commands...
Страница 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Страница 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Страница 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Страница 260: ...15 22 Additional SCPI Commands...
Страница 261: ...A Specifications Specifications...
Страница 263: ...B Statusand ErrorMessages StatusandError Messages...
Страница 268: ...B 6 Status and Error Messages...
Страница 269: ...C Measurement Considerations Measurement Consider ations...
Страница 278: ...C 10 Measurement Considerations...
Страница 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Страница 284: ...D 6 Model 182 Emulation Commands...
Страница 285: ...E Example Programs ExamplePro grams...
Страница 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Страница 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Страница 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Страница 311: ...H Measurement Queries Measurement Queries...
Страница 316: ...H 6 Measurement Queries...
Страница 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...