Disassembly
5-3
Handling and cleaning
To avoid contaminating PC board traces with body oil or other foreign matter, avoid touching
the PC board traces while you are repairing the instrument. Motherboard areas covered by the
shield have high-impedance devices or sensitive circuitry where contamination could cause
degraded performance.
Handling PC boards
Observe the following precautions when handling PC boards:
•
Wear cotton gloves.
•
Only handle PC boards by the edges and shields.
•
Do not touch any board traces or components not associated with repair.
•
Do not touch areas adjacent to electrical contacts.
•
Use dry nitrogen gas to clean dust off PC boards.
Solder repairs
Observe the following precautions when soldering a circuit board:
•
Use an OA-based (organic activated) flux, and take care not to spread the flux to other
areas of the circuit board.
•
Remove the flux from the work area when you have finished the repair by using pure
water with clean, foam-tipped swabs or a clean, soft brush.
•
Once you have removed the flux, swab only the repair area with methanol, then blow dry
the board with dry nitrogen gas.
•
After cleaning, allow the board to dry in a 50°C, low-humidity environment for several
hours.
Содержание 2015
Страница 15: ...1 Performance Verification...
Страница 34: ...2 Calibration...
Страница 60: ...3 Routine Maintenance...
Страница 64: ...4 Troubleshooting...
Страница 89: ...5 Disassembly...
Страница 100: ...5 12 Disassembly...
Страница 101: ...6 Replaceable Parts...
Страница 115: ...A Specifications...
Страница 127: ...B Calibration Reference...
Страница 149: ...C Calibration Program...
Страница 158: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...