Disassembly
Disassembly procedures
Case cover removal
Follow the steps below to remove the case cover to gain access to internal parts.
WARNING
Before removing the case cover, disconnect the line cord and any test leads
from the instrument.
1.
Remove Handle — The handle serves as an adjustable tilt-bail. Adjust its position by
gently pulling it away from the sides of the instrument case and swinging it up or down.
To remove the handle, swing the handle below the bottom surface of the case and back
until the orientation arrows on the handles line up with the orientation arrows on the
mounting ears. With the arrows lined up, pull the ends of the handle away from the case.
2.
Remove Mounting Ears — Remove the screw that secures each mounting ear. Pull down
and out on each mounting ear.
NOTE
When re-installing the mounting ears, make sure to mount the right ear to the right
side of the chassis, and the left ear to the left side of the chassis. Each ear is marked
“RIGHT” or “LEFT” on its inside surface.
3.
Remove Rear Bezel — To remove the rear bezel, loosen the two captive screws that se-
cure the rear bezel to the chassis. Pull the bezel away from the case.
4.
Removing Grounding Screws — Remove the two grounding screws that secure the case
to the chassis. They are located on the bottom of the case at the back.
5.
Remove Cover — To remove the case, grasp the front bezel of the instrument, and care-
fully slide the chassis forward. Slide the chassis out of the metal case.
NOTE
To gain access to the components under the DMM board shield, remove the shield,
which is secured to the DMM board by a single screw.
DMM board removal
Perform the following steps to remove the DMM (106) board. This procedure assumes that
the case cover is already removed.
1.
Remove the IEEE-488 and RS-232 fasteners.
The IEEE-488 and the RS-232 connectors each have two nuts that secure the connectors
to the rear panel. Remove these nuts.
2.
Remove the front/rear switch rod.
At the switch, place the edge of a flat-blade screw driver in the notch on the pushrod.
Gently twist the screw driver while pulling the rod from the shaft.
Содержание 2015
Страница 15: ...1 Performance Verification...
Страница 34: ...2 Calibration...
Страница 60: ...3 Routine Maintenance...
Страница 64: ...4 Troubleshooting...
Страница 89: ...5 Disassembly...
Страница 100: ...5 12 Disassembly...
Страница 101: ...6 Replaceable Parts...
Страница 115: ...A Specifications...
Страница 127: ...B Calibration Reference...
Страница 149: ...C Calibration Program...
Страница 158: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...