B-12
Calibration Reference
:STEP8
(:CALibration:PROTected:DC:STEP8)
Purpose
To program the 100k
Ω
4-wire comprehensive calibration step.
Format
:cal:prot:dc:step8 <Cal_resistance>
Parameter
<Cal_resistance> = 90E3 to 110E3 [
Ω
]
Description
:STEP8
programs the 100k
Ω
4-wire resistance comprehensive calibration
step. The allowable range of the calibration resistance parameter is from
90E3 to 110E3, but 100E3 is recommended for best results.
Example
:CAL:PROT:DC:STEP8 100E3
Program 100k
Ω
step.
:STEP9
(CALibration:PROTected:DC:STEP9)
Purpose
To program the 1M
Ω
comprehensive calibration step.
Format
:cal:prot:dc:step9 <Cal_resistance>
Parameter
<Cal_resistance> = 900E3 to 1.1E6 [
Ω
]
Description
:STEP9
programs the 1M
Ω
comprehensive calibration step. The allowable
range of the calibration resistance parameter is from 900E3 to 1.1E6. Use the
1E6 value whenever possible, or the closest possible value.
Example
:CAL:PROT:DC:STEP9
Program 1M
Ω
calibration step.
:STEP10
(CALibration:PROTected:DC:STEP10)
Purpose
To program the 10mA comprehensive calibration step.
Format
:cal:prot:dc:step10 <Cal_current>
Parameter
<Cal_current> = 9E-3 to 11E-3 [A]
Description
:STEP10
programs the 10mA comprehensive calibration step. The allow-
able range of the calibration current parameter is from 9E-3 to 11E-3. Use
the 10E-3 value whenever possible for best results.
Example
:CAL:PROT:DC:STEP10 10E-3
Program 10mA step.
Содержание 2015
Страница 15: ...1 Performance Verification...
Страница 34: ...2 Calibration...
Страница 60: ...3 Routine Maintenance...
Страница 64: ...4 Troubleshooting...
Страница 89: ...5 Disassembly...
Страница 100: ...5 12 Disassembly...
Страница 101: ...6 Replaceable Parts...
Страница 115: ...A Specifications...
Страница 127: ...B Calibration Reference...
Страница 149: ...C Calibration Program...
Страница 158: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...