B-10
Calibration Reference
:STEP2
(:CALibration:PROTected:DC:STEP2)
Purpose
To perform front terminal open-circuit calibration.
Format
:cal:prot:dc:step2
Parameter
None
Description
:STEP2
performs the open-circuit calibration step in the comprehensive cal-
ibration procedure. Disconnect all cables and accessories from the input
jacks before sending this command.
Example
:CAL:PROT:DC:STEP2
Perform open circuit calibration.
:STEP3
(:CALibration:PROTected:DC:STEP3)
Purpose
To program the +10V comprehensive calibration step.
Format
:cal:prot:dc:step3 <Cal_voltage>
Parameter
<Cal_voltage> = 9 to 11 [V]
Description
:STEP3
programs the +10V DC comprehensive calibration step. The allow-
able range of the calibration voltage parameter is from 9 to 11, but 10 is rec-
ommended for best results.
Example
:CAL:PROT:DC:STEP3 10
Program 10V step.
:STEP4
(:CALibration:PROTected:DC:STEP4)
Purpose
To program the -10V DC comprehensive calibration step.
Format
:cal:prot:dc:step4 <Cal_voltage>
Parameter
<Cal_voltage> = -9 to -11 [V]
Description
:STEP4
programs the -10V DC comprehensive calibration step. The allow-
able range of the calibration voltage parameter is from -9 to -11, but -10 is
recommended for best results.
Example
:CAL:PROT:DC:STEP4 -10
Program -10V step.
Содержание 2015
Страница 15: ...1 Performance Verification...
Страница 34: ...2 Calibration...
Страница 60: ...3 Routine Maintenance...
Страница 64: ...4 Troubleshooting...
Страница 89: ...5 Disassembly...
Страница 100: ...5 12 Disassembly...
Страница 101: ...6 Replaceable Parts...
Страница 115: ...A Specifications...
Страница 127: ...B Calibration Reference...
Страница 149: ...C Calibration Program...
Страница 158: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...