Performance Verification
1-17
Verifying frequency
Follow the steps below to verify the Model 2015 frequency function:
1.
Connect the frequency synthesizer to the Model 2015 INPUT jacks. (See Figure 1-7.)
2.
Set the synthesizer to output a 1kHz, 1V RMS sine wave.
3.
Select the Model 2015 frequency function by pressing the FREQ key.
4.
Verify that the Model 2015 frequency reading is between 0.9999kHz and 1.0001kHz.
Figure 1-7
Connections for frequency verification
Model 2015
2015 THD MULTIMETER
RANGE
!
F
500V
PEAK
FRONT/REAR
3A 250V
AMPS
HI
INPUT
LO
SENSE
Ω
4 WIRE
INPUTS
350V
PEAK
1000V
PEAK
AUTO
SHIFT
LOCAL
POWER
RANGE
R
SHIFT
CH1
REM
TALK
LSTN
SRQ
STAT
REL
FILT
4W
BUFFER
MATH
REAR
SCAN
TIMER
STEP
CH2
CH3
CH4
CH5
CH6
CH7
CH8
CH9
CH10
HOLD
TRIG
FAST
MED
SLOW
AUTO
ERR
EXIT
ENTER
DIGITS RATE
REL
FILTER
TRIG
EX TRIG
STORE RECALL
SOURCE MEAS
DCV
DCI
MATH
T H D
dBm
ACV
ACI
Ω
2
Ω
4
FREQ
TEMP
dB
CONT
PERIOD TCOUPL
LIMITS
ON/OFF
DELAY
HOLD
SAVE
SETUP
CONFIG
HALT
TEST
RS232
GPIB
CAL
STEP
SCAN
BNC-to-Dual
Banana Plug
Adapter
50
Ω
BNC Coaxial Cable
Frequency Synthesizer
Main
Function
Output
THD
Содержание 2015
Страница 15: ...1 Performance Verification...
Страница 34: ...2 Calibration...
Страница 60: ...3 Routine Maintenance...
Страница 64: ...4 Troubleshooting...
Страница 89: ...5 Disassembly...
Страница 100: ...5 12 Disassembly...
Страница 101: ...6 Replaceable Parts...
Страница 115: ...A Specifications...
Страница 127: ...B Calibration Reference...
Страница 149: ...C Calibration Program...
Страница 158: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...