Performance Verification
1-15
6.
Source the nominal full-scale resistance values for the 100
Ω
-10M
Ω
ranges summarized
in Table 1-6, and verify that the readings are within calculated limits.
7.
Connect the Model 2015 INPUT and SENSE jacks to the calibrator as shown in
Figure 1-6.
8.
Disable external sense on the calibrator.
9.
Set the Model 2015 for the 100M
Ω
range.
10.
Source a nominal 100M
Ω
resistance value, and verify that the reading is within cal-
culated limits for the 100M
Ω
range.
Figure 1-6
Connections for resistance verification (100M
Ω
range)
Table 1-6
Limits for resistance verification
Ω
Range
Nominal
resistance
Nominal reading limits
(1 year, 18°C-28°C)
Recalculated limits*
100
Ω
1k
Ω
10k
Ω
100k
Ω
1M
Ω
10M
Ω
100M
Ω
100
Ω
1k
Ω
10k
Ω
100k
Ω
1M
Ω
10M
Ω
100M
Ω
99.9860 to 100.0140
Ω
0.999890 to 1.000110k
Ω
9.99890 to 10.00110k
Ω
99.9890 to 100.0110k
Ω
0.999890 to 1.000110M
Ω
9.99590 to 10.00410M
Ω
99.8470 to 100.1530M
Ω
__________ to __________
Ω
__________ to __________ k
Ω
__________ to __________ k
Ω
__________ to __________ k
Ω
__________ to __________ M
Ω
__________ to __________ M
Ω
__________ to __________ M
Ω
* Calculate limits based on actual calibration resistance values and Model 2015 one-year accuracy
specifications. See Verification limits.
Output
LO
2015 THD MULTIMETER
RANGE
!
F
500V
PEAK
FRONT/REAR
3A 250V
AMPS
HI
INPUT
LO
SENSE
Ω
4 WIRE
INPUTS
350V
PEAK
1000V
PEAK
AUTO
SHIFT
LOCAL
POWER
RANGE
R
SHIFT
CH1
REM
TALK
LSTN
SRQ
STAT
REL
FILT
4W
BUFFER
MATH
REAR
SCAN
TIMER
STEP
CH2
CH3
CH4
CH5
CH6
CH7
CH8
CH9
CH10
HOLD
TRIG
FAST
MED
SLOW
AUTO
ERR
EXIT
ENTER
DIGITS RATE
REL
FILTER
TRIG
EX TRIG
STORE RECALL
SOURCE MEAS
DCV
DCI
MATH
T H D
dBm
ACV
ACI
Ω
2
Ω
4
FREQ
TEMP
dB
CONT
PERIOD TCOUPL
LIMITS
ON/OFF
DELAY
HOLD
SAVE
SETUP
CONFIG
HALT
TEST
RS232
GPIB
CAL
STEP
SCAN
Model 2015
Input
LO
Input
HI
Resistance Calibrator
Note: Use shielded cables to minimize noise.
Disable calibrator external sense mode.
Sense HI
Sense LO
Output
HI
THD
Содержание 2015
Страница 15: ...1 Performance Verification...
Страница 34: ...2 Calibration...
Страница 60: ...3 Routine Maintenance...
Страница 64: ...4 Troubleshooting...
Страница 89: ...5 Disassembly...
Страница 100: ...5 12 Disassembly...
Страница 101: ...6 Replaceable Parts...
Страница 115: ...A Specifications...
Страница 127: ...B Calibration Reference...
Страница 149: ...C Calibration Program...
Страница 158: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...