Performance Verification
1-11
2.
Select the AC volts function by pressing the
ACV
key, then choose the SLOW integra-
tion rate with the RATE key.
3.
Set the Model 2015 for the 100mV range; make sure that REL is disabled.
4.
Source 1kHz and 50kHz AC voltages for each of the ranges summarized in Table
1-3, and make sure that the respective Model 2015 readings fall within stated limits.
Table 1-3
ACV reading limits
ACV
range
Applied AC
voltage
1kHz reading limits
(1 year, 18°C-28°C)
50kHz reading limits
(1 year, 18°C-28°C)
100mV
1V
10V
100V
750V
100.0000mV
1.000000V
10.00000V
100.0000V
700.000V*
99.910 to 100.090mV
0.99910 to 1.00090V
9.9910 to 10.0090V
99.910 to 100.090V
699.36 to 700.64V
99.830 to 100.170mV
0.99830 to 1.00170V
9.98300 to 10.0170V
99.830 to 100.170V
698.79 to 701.21V
* If the 5725A amplifier is not available, change the 700V @ 50kHz step to 220V @
50kHz. Reading limits for 220V @ 50kHz = 219.36 to 220.64V.