1-12
Performance Verification
Verifying DC current
Check DC current accuracy by applying accurate DC currents from the DC current calibrator
to the AMPS input of the Model 2015 and verifying that the displayed readings fall within spec-
ified limits.
Follow these steps to verify DC current accuracy:
1.
Connect the Model 2015 AMPS and INPUT LO jacks to the calibrator as shown in
Figure 1-3.
Figure 1-3
Connections for DC current verification
2.
Select the DC current measurement function by pressing the
DCI
key, then choose the
SLOW integration rate with the RATE key.
3.
Set the Model 2015 for the 10mA range.
4.
Source positive and negative full-scale currents for each of the ranges listed in Table
1-4, and verify that the readings for each range are within stated limits.
Table 1-4
DCI limits
DCI
range
Applied DC current*
Reading limits (1 year, 18°C-28°C)
10mA
100mA
1A
3A
10.0000mA
100.0000mA
1.000000A
2.20000A
9
.99460 to 10.00540mA
99.9100 to 100.0900mA
0.999160 to 1.000840A
2.197315 to 2.202685A
*Source positive and negative currents with values shown.
Output HI
Output
LO
2015 THD MULTIMETER
RANGE
!
F
500V
PEAK
FRONT/REAR
3A 250V
AMPS
HI
INPUT
LO
SENSE
Ω
4 WIRE
INPUTS
350V
PEAK
1000V
PEAK
AUTO
SHIFT
LOCAL
POWER
RANGE
R
SHIFT
CH1
REM
TALK
LSTN
SRQ
STAT
REL
FILT
4W
BUFFER
MATH
REAR
SCAN
TIMER
STEP
CH2
CH3
CH4
CH5
CH6
CH7
CH8
CH9
CH10
HOLD
TRIG
FAST
MED
SLOW
AUTO
ERR
EXIT
ENTER
DIGITS RATE
REL
FILTER
TRIG
EX TRIG
STORE RECALL
SOURCE MEAS
DCV
DCI
MATH
T H D
dBm
ACV
ACI
Ω
2
Ω
4
FREQ
TEMP
dB
CONT
PERIOD TCOUPL
LIMITS
ON/OFF
DELAY
HOLD
SAVE
SETUP
CONFIG
HALT
TEST
RS232
GPIB
CAL
STEP
SCAN
Model 2015
Input
LO
Note: Be sure calibrator is set for
Amps
DC Current Calibrator
normal current output.
THD
Содержание 2015
Страница 15: ...1 Performance Verification...
Страница 34: ...2 Calibration...
Страница 60: ...3 Routine Maintenance...
Страница 64: ...4 Troubleshooting...
Страница 89: ...5 Disassembly...
Страница 100: ...5 12 Disassembly...
Страница 101: ...6 Replaceable Parts...
Страница 115: ...A Specifications...
Страница 127: ...B Calibration Reference...
Страница 149: ...C Calibration Program...
Страница 158: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...