9 Installation
4.
Select
Drive to packing position
.
NOTICE!
The device supports driving to Packing position in cases of Z-
movement failure.
5.
Press and hold down the exposure switch to drive the unit to the correct height.
6.
Power the device off.
7.
Detach the device from mains power and detach any external cables.
8.
Detach the covers in reverse order to what is described in chapter
Attaching the
covers
.
NOTICE!
It's recommended to wrap protective material around the covers
and place them in boxes to protect them against scratches.
9.
Detach the column interconnection cables, install the packaging supports back to
their original place and install the plywood supports in reverse order to what is
described in chapter
Installing the device
.
10.
Loosen the three locking screws holding the carriage on the column and turn them
outwards.
11.
Carefully push the latch inside the carriage with a flat tool to release the carriage and
carefully lower it to the floor.
NOTICE!
This step requires at least two persons.
NOTICE!
It is recommended to place packing foam or such on the floor to
prevent any damage to the carriage.
12.
Detach the carriage from the column, detach the column from the wall and floor and
move the carriage and the column onto the transportation pallet.