9 Installation
9.9.10 Verification
The device offers Panoramic and 3D verification test programs with the device
movements disabled. These programs can be used for radiation related measurements,
for example, during device installation and acceptance tests.
A separate radiation measuring device/equipment is required for taking the radiation
measurements.
To access the verification tests:
1.
Go to Settings menu.
2.
Select
Service menu
and enter PIN code
0612
.
3.
Select
Verification
menu that appears under the Service menu.
4.
Select the
PAN testing
tab for Panoramic verification test or
3D testing
tab for 3D
verification test.
The
PAN Testing
program allows adjustment of the exposure values
kV
and
mA
for the
tests.
The
3D Testing
program allows selection of different radiaton modes High/ENDO
Resolution (Constant), Standard Resolution (Pulse 10ms) or LDT resolution (Pulse 5ms)
and adjusting the related
mA
.
NOTICE!
When using the
PAN testing
program, the X-Ray beam is adjusted
wider (over 2 cm) than in normal imaging programs to allow easier positioning of
the measurement device.
9.9.11 Repacking the device
If the device needs to be re-located it must be dismantled and is advisable to be
repacked for safer and easier transportation.
1. GUI:
Go to Settings menu.
2. GUI:
Select
Service menu
and enter PIN code
0612
.
3. GUI:
Select
Service
menu that appears under the Service menu.
96