8 Troubleshooting
8 Troubleshooting
Problem
Possible cause
Solution
Taken image is not
transferred to the modality
workstation.
Local network connection
is disrupted, which causes
loss of data.
The device stores the latest
image until a confirmation
of a succesfull transfer to
the workstation has been
received. Re-establish the
local network connection
and the image data is
transferred automatically.
Do not power the device off
or the image will be erased.
Device cannot be driven
upwards or downwards.
1.
Up/down slider is on
safe mode.
2.
Mechanical safety
device on the driving
motor has activated.
3.
Positioning panel not
working.
1.
Clean the positioning
panel and wait for the
safe mode to clear out.
2.
Contact service.
No connection to the device.
1.
The device is not
powered on.
2.
Problem with the local
area network (LAN).
3.
Check the device
configuration.
1.
Power the device on.
2.
Check the Ethernet
cable connections on
the workstation and the
device.
3.
Restart the device and
the workstation.
ORTHOPANTOMOGRAPH
™
OP 3D
53