9 Installation
2.
The positioning lights are held in place in a fixture with one screw. Loosen the screw
to release the light.
Horizontal lights
Midsagittal light
3.
Re-align the light by rotating it and tilting it up/down.
4.
Tighten the screw.
5.
Re-attach the covers.
NOTICE!
When reattaching the left side cover, ensure that the horizontal
lasers sit in the middle of the groove so that the cover does not block the
positioning light beam
6.
Attach the 3D Geometry Calibration Phantom to the device and verify the correct
alignment of the lights with Device Geometry calibration program selected.
7.
Repeat the process if lights are still not aligned correctly.
92
ORTHOPANTOMOGRAPH
™
OP 3D