9 Installation
9.7 Finishing the installation
After the device has been installed, perform these checks and tasks, before handing it to
the user.
1.
Check the wall mount screw tightness.
2.
Check the floor screw tightness.
3.
Ensure that the wall mount screws match the wall material.
4.
Ensure that the wall mount and floor screws withstand 1500N pull-out force.
5.
Check that the power cord and the Ethernet cable, along with all the other cables
connected, are installed so that the up/down movement of the device doesn't cause
stress to them.
6.
Check that the power cord is not damaged in any way
7.
Check that all covers are attached correctly.
8.
Check the electrical safety of the device by inspecting the proper grounding of user
touchable metallic parts.
9.
Check that the IEC 60601-1 leakage current levels are not exceeded.
10.
Check the functionality of the power switch.
11.
Check that the patient positioning lights work properly and are aligned correctly. The
positioning light can be checked during the Device Geometry Calibration.
12.
Check that the device has been configured correctly for the installation location. See
chapter
Device configuration
for more information.
13.
Carry out the device calibrations described in chapter
Complete device calibration
.
14.
Check that the exposure warning indicators work correctly.
15.
Check the functionality of the Emergency Stop button.
NOTICE!
Activating the Emergency Stop causes the device to restart.
16.
Check the Z-movement limits.
17.
Gather the screws left over from the installation for storage.