9 Installation
32.
Detach the two screws holding the front support bar in place, push it inside the
upper shelf frame and reattach it.
33.
Detach the linear movement support from the top shelf (2 screws).
34.
Detach the rotating unit support from the carriage (4 screws), push the rotating unit
outwards a bit and lift the support off from the rotating unit .
NOTICE!
The rotating unit can now move freely, ensure that it does not hit
the lower shelf.
35.
Connect the Column control and Emergency switch connector to the connectors in
the column.