9 Installation
9.
The QC image appears on the workstation, showing the result of the check.
10. Workstation:
Visually evaluate the image using the dental imaging software for
visual defects such as artifacts.
NOTICE!
The device determines if the QC image is PASSED or FAILED
according to measured data, not based on the visible image quality.
NOTICE!
You should also compare the new QC image to the reference image
taken during the installation or the latest service. Doing this helps you ensure
that the image quality has remained constant.
11.
If the image fails or any visual defects are apparent, redo the QC program. If it fails
again, recalibrate the device or contact service.
12.
Save the passed QC image as a reference image for the user and let the user know
the reference image location.
82
ORTHOPANTOMOGRAPH
™
OP 3D