9 Installation
9.8.2 Device Movement Calibration
This program is used to calibrate the movements of the upper shelf and the rotating unit.
No calibration tools are required when performing this calibration.
1.
Select
Device Movement Calibration
from the Calibrations menu.
2.
Turn the device to HOME (Patient-in) position by hand.
3.
Press and hold the exposure button. The device starts to move.
4.
Keep the exposure button pressed down during the whole calibration.
5.
When the device's movements stop and the program end tone is played, the
program is complete.
9.8.3 Z-Movement Limits
This program is used to determine the maximum height of the device, required especially
for low room installations.
No calibration tools are required when performing this calibration.
NOTICE!
The maximum height of the device must be at least 2080 mm / 81.9".
NOTICE!
Z-movement limits do not need to be re-determined during annual
calibration, checking the limits will suffice.
1.
Select
Z-Movement Limits
from the Calibrations menu.
2.
Press and hold down the exposure button. The carriage starts to move upwards.
3.
Keep the exposure button pressed down until the device is in the maximum allowed
height and release the button.
4.
Wait a minimum of 2 seconds and then Press and hold the exposure button down to
drive the device downwards.
5.
When the device's movements stop and the program end tone is played, the
program is complete.