5. S2terminal and service commands
ORTHOPANTOMOGAPH OP 2D
87
NOTICE!
The "
still
" command, described later, generates x-rays
and activates the CCD sensor but the rotating unit
does not move.
expcnt
exposure counter, use [<clear>] option to clear it
Displays the total number of exposures taken with the
tubehead.
To zero the exposure counter enter clear,
____________________________________________
>expcnt clear
____________________________________________
and then press
Enter
.
NOTICE!
Always zero the exposure counter when the tubehead
is replaced.
help
display info on command (‘help [cmd]’ or ‘h [cmd]’)
Displays the service command list and information
about the commands.
imagecalib takes a calibration image for calibrating the DSD
Creates a technical image which is then used by the im-
age capturing driver for image correction.
This function can also be carried out using the unit GUI.
See the section "Calibrating the CCD sensor" in the
installation manual.
ip
con
fi
gure ip address
Displays the current IP address of the unit.
To change the IP address enter the new one,
____________________________________________
> ip
aaa.bbb.ccc.ddd (the new IP address)
____________________________________________
and then press
Enter
.
Содержание ORTHOPANTOMOGAPH OP 2D
Страница 1: ...ORTHOPANTOMOGAPH OP 2D Digital Panoramic X ray Unit Service Manual ENGLISH 216143 rev 1 0 805 5058...
Страница 2: ......
Страница 4: ...IV ORTHOPANTOMOGAPH OP 2D Contents...
Страница 24: ...14 ORTHOPANTOMOGAPH OP 2D 2 Unit Description 2 3 Unit dimensions...
Страница 32: ...22 ORTHOPANTOMOGAPH OP 2D 2 Unit Description...
Страница 33: ...2 Unit Description ORTHOPANTOMOGAPH OP 2D 23 2 6 Block diagram...
Страница 34: ...24 ORTHOPANTOMOGAPH OP 2D 2 Unit Description 2 7 Fuse diagram...
Страница 38: ...28 ORTHOPANTOMOGAPH OP 2D 2 Unit Description...
Страница 43: ...3 Circuit Boards L1200 ORTHOPANTOMOGAPH OP 2D 33 L1200 Block Diagram...
Страница 46: ...36 ORTHOPANTOMOGAPH OP 2D 3 Circuit Boards L1500 L1500 Block diagram...
Страница 50: ...40 ORTHOPANTOMOGAPH OP 2D 3 Circuit Board L1600 L1600 Block diagram...
Страница 54: ...44 ORTHOPANTOMOGAPH OP 2D 3 Circuit Boards L1700 L1700 Block Diagram...
Страница 57: ...3 Circuit Boards L1800 ORTHOPANTOMOGAPH OP 2D 47 L1800 Block diagram...
Страница 61: ...3 Circuit Boards P1910 ORTHOPANTOMOGAPH OP 2D 51 P1910 Block diagram...
Страница 64: ...54 ORTHOPANTOMOGAPH OP 2D 3 Circuit Boards L2000 L2000 Indicator LEDs None L2000 Test Points None...
Страница 66: ...56 ORTHOPANTOMOGAPH OP 2D 3 Circuit Boards L2100 L2100 Indicator LEDs None L2100 Test Points None...