86
ORTHOPANTOMOGAPH
OP
2D
5. S2terminal and service commands
endu
endurance test, drives rotating unit 75 rounds
Automatically drives the rotating unit through 75 rotation
cycles. No radiation in generated during rotation.
This command is used during factory testing.
exitdemo
exit demo mode
Exits the unit demonstration mode.
exp
exposure (<time>) in milliseconds, default is 1000
Generates x-rays. The CCD sensor is not activated and
the rotating unit does not move.
It is used to check the beam alignment during unit
installation (refer to the Installation manual for more
information), and to trouble shoot the tubehead and
generator board.
CAUTION:
Protect yourself from radiation when using this
function.
The default exposure time is 1000 ms (when "exp" is
keyed in). If you wish to have a longer exposure time
key in "exp" and them enter a value, for example:
____________________________________________
>exp 3000
____________________________________________
will result an exposure time of 3000 ms. The minimum
value you can key in is 100 ms and the maximum is
20000.
Press and hold down the exposure button to activate
x-rays.
If exposure times of 1000 ms or longer are used, the
mAfb and kVfb ADC values will appear.
To exit the command key in
q
and then press
Enter
or
press the
RETURN
key on the unit control panel.
NOTICE!
The kV value used for the exposure will be the one
selected before entering the service mode. The kV
value cannot be changed while the unit is in the service
mode. If you wish to change the kV, you must exit the
service mode.
Содержание ORTHOPANTOMOGAPH OP 2D
Страница 1: ...ORTHOPANTOMOGAPH OP 2D Digital Panoramic X ray Unit Service Manual ENGLISH 216143 rev 1 0 805 5058...
Страница 2: ......
Страница 4: ...IV ORTHOPANTOMOGAPH OP 2D Contents...
Страница 24: ...14 ORTHOPANTOMOGAPH OP 2D 2 Unit Description 2 3 Unit dimensions...
Страница 32: ...22 ORTHOPANTOMOGAPH OP 2D 2 Unit Description...
Страница 33: ...2 Unit Description ORTHOPANTOMOGAPH OP 2D 23 2 6 Block diagram...
Страница 34: ...24 ORTHOPANTOMOGAPH OP 2D 2 Unit Description 2 7 Fuse diagram...
Страница 38: ...28 ORTHOPANTOMOGAPH OP 2D 2 Unit Description...
Страница 43: ...3 Circuit Boards L1200 ORTHOPANTOMOGAPH OP 2D 33 L1200 Block Diagram...
Страница 46: ...36 ORTHOPANTOMOGAPH OP 2D 3 Circuit Boards L1500 L1500 Block diagram...
Страница 50: ...40 ORTHOPANTOMOGAPH OP 2D 3 Circuit Board L1600 L1600 Block diagram...
Страница 54: ...44 ORTHOPANTOMOGAPH OP 2D 3 Circuit Boards L1700 L1700 Block Diagram...
Страница 57: ...3 Circuit Boards L1800 ORTHOPANTOMOGAPH OP 2D 47 L1800 Block diagram...
Страница 61: ...3 Circuit Boards P1910 ORTHOPANTOMOGAPH OP 2D 51 P1910 Block diagram...
Страница 64: ...54 ORTHOPANTOMOGAPH OP 2D 3 Circuit Boards L2000 L2000 Indicator LEDs None L2000 Test Points None...
Страница 66: ...56 ORTHOPANTOMOGAPH OP 2D 3 Circuit Boards L2100 L2100 Indicator LEDs None L2100 Test Points None...