Header
Message text
Argon-Beamer Fault
Please check if the argon bottles are connected and open. Empty bottles should be replaced. Subsequently restart the Argon-
Beamer by activating the flashing ‘Purge’ button.
TSI Argon-Beamer
Warning
The annual Technical Safety Inspection (TSI) for the Argon-Beamer is due.
Coded System
Notice
The lifetime of the instrument is ending soon. Please order a replacement in good time. Any use of the instrument beyond its lifetime
is not covered by warranty. Please contact your dealer in good time to purchase a new instrument.
Coded System
Warning
The maximum lifetime of the instrument has been reached. Any further use is not covered by warranty.
The maximum service lifetime of the instruments must not be exceeded, in order to guarantee safe usage. Any further use is at the
user's risk.
Coded System
Warning
A software update is necessary to use the Coded System with this instrument.
Only carry out manual settings at this instrument.
Please contact the Technical Support.
Contact: MENU – SERVICE.
Coded System
Warning
The preferred parameters for this coded instrument cannot be loaded.
Only carry out manual settings at this instrument.
Please contact the Technical Support.
Contact: MENU – SERVICE.
Internal Error xxxx
(e.g. with
xxxx=4183)
If this message appears again, please contact the Technical Support. Contact: MENU – SERVICE.
With internal errors a number is shown in the
header.
Please quote this number to the Technical
Support.