223
EMV
CEM
Guidance and manufacturer’s declaration – electromagnetic emissions (IEC 60601-1-2, Table 1)
AUTOCON
®
III 400 is intended for use in the electromagnetic environment specified below. The customer or user of the AUTOCON
®
III 400 must make sure
that it is used in such an environment.
Interference emissions measurements
Compliance
Electromagnetic environment – guidelines
HF emissions acc. to CISPR 11
Group 2
The AUTOCON
®
III 400 must emit electromagnetic
energy in order to perform its intended function.
Nearby electronic devices may be affected.
HF emissions acc. to CISPR 11
Class B
The AUTOCON
®
III 400 is suitable for use in all
establishments including domestic establishments
and those directly connected to the public low-
voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions acc. to IEC 61000-3-2
Class A
Voltage fluctuations/flicker emissions acc. to
IEC 61000-3-3
Complies
EMC